Section Iv - Required Equipment; Section V - System Setup - Tyco KENDALL GENIUS 2 Manual De Operación Y Servicio

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 55
Status: Approved Doc#: AW10001726 Effective Date: nulldate

Section IV - Required Equipment

1. GENIUS 2 Checker/Calibrator
2. Power module with AC adapter
3. Probe cable
4. Operator's Manual
5. GENIUS 2 Thermometers to be tested
6. GENIUS 2 disposable probe covers
7. Small Phillips screwdriver

Section V - System Setup

The following actions should precede use of the GENIUS 2 Checker/Calibrator:
1. Connect the correct AC plug adapter for your country to the power module.
2. Connect the power module cable to the power jack on the front panel.
3. Connect the power module to an AC power outlet.
4. Turn the unit on via the switch on the front panel.
5. Let the GENIUS 2 Checker/Calibrator unit warm up for at least 15 minutes.
6. Set aside a cassette with new GENIUS 2 probe covers.
7. Let the GENIUS 2 Thermometers equilibrate to the room temperature.
8. Prepare the GENIUS 2 Thermometers by thoroughly cleaning the lenses prior to use. Refer to the GENIUS 2 Thermometer
operating manual for cleaning instructions.
9. Remove the battery compartment access panel by using a small Phillips screwdriver. Remove all three AAA batteries.
10. Connect the probe cable round connector to the front panel of the GENIUS 2 Checker/Calibrator.
11. Connect the other end of the probe connector into the battery compartment. The connector will snap into place where the
batteries were. Refer to Figure 2.
Figure 2 - Insertion of Probe Connector into GENIUS 2 Battery Compartment
GENIUS ™
2
Checker / Calibrator
4
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
Operation and Service Manual

Publicidad

Tabla de contenido
loading

Productos relacionados para Tyco KENDALL GENIUS 2

Tabla de contenido