Figure 3 Remove the sensor

8.1.2 Replace the membrane

Refer to the steps that follow and
Figure 4
to replace the sensor membrane.
1. Unscrew the filling screw and remove along with the gasket.
2. Unscrew the holding nut.
3. Carefully remove the electrode from the sensor body.
4. Empty any electrolyte left in the sensor body.
5. Unscrew the old membrane from the sensor body.
6. Install the new membrane. Refer to
Assemble the sensor
on page 6.
English 15