TCM 3000 BL
Fault:
Device does not
function at all:
à à à à Display not
illuminated
Footpedal does not
function:
à à à à Display illuminated
Motor does not run:
à à à à Display illuminated
If a fault cannot be eliminated, please contact the supplier or an authorised service agent.
Description
Electronic motor ENT ...................................................................................................................2037nou
Conform cannula handpiece ..............................................................................................................5074
One-way hose set for Conform cannula handpiece ........................................................................28966
For orders of other parts, please contact our customer service agents who will be glad to assist
you.
8
8 Faults and fault detection
Cause:
Control unit not switched
on
No connection to the
mains
Defective fuse
Wrong operating voltage
Footpedal not connected
Incorrect operation
No motor connection
Footpedal step not
pressed down
Incorrect operation
9 Spare parts with order numbers
Elimination:
Switch on main switch
"I/O" at the back of unit to
"I" position.
Connect control unit to
power supply.
Check fuses and replace if
necessary
Check mains voltage
Plug in footpedal on the
front side of the control
unit
Check up operating
instructions
Plug in motor cable on the
front side of control unit
Press down step
Check up operating
instructions
NOUVAG AG
Reference in
operating
instructions:
5.1. Switching on the TCM
3000 BL
4.1 Connection to power
supply
7.1 Exchanging control unit
fuses
1.2 Technical data
7.1 Exchanging control unit
fuses
4.2 Footpedal connection
4.3 Motor connection
5.4 Start motor
Art. No.
No. 31739 35/05