OPeratInG the UnIt
calibrating the thermometer
This function is only available in the "BoDY" measuring mode. This function is designed to
assist you in adjusting the Promed Irt-60 to the measuring results of your usual clinical
thermometer in order to achieve comparable measured values if you have always measured
your body temperature using a thermometer rectally, but find this unpleasant. You can
calibrate the Promed Irt-60 by up to ±3.0° c (±5.4° F).
In order to perform the calibration, press the button (F) and keep it pressed for two (2)
seconds. Then press the button (F) three (3) times.
You can then calibrate the thermometer as follows:
Pressing the measuring button (D) increases the value by 0.1° c.
Pressing the memory button (e) reduces the value by 0.1° c.
In order to save the set values, press the button (F) again.
measuring the room temperature
In order to measure the room temperature, you must switch the thermometer to the "Room"
measuring mode. To do this, press the button (F) and keep it pressed for two (2) seconds.
Press then the button (F) another two (2) times.
By pressing the measuring button (D) or by pressing the memory button (e), you can switch
the measuring mode:
In order to save the set measuring mode, press the button (F) another two (2) times.
In order to perform a measurement, switch on the thermometer as described above. Then
press the measuring button (D) once. You hear a short beep and can read the measuring
result from the display (b).
Please note:
The thermometer must have been in the room for at least 10 minutes before the measurement
can be carried out.
measuring surface temperatures
In order to measure the surface temperature, you must switch the thermometer to the
"Surface temp" measuring mode. For this purpose, press the button (F) and keep it pressed
for two (2) seconds. Press then the button (F) another two (2) times.
By pressing the measuring button (D) or by pressing the memory button (e), you can switch
the measuring mode:
In order to save the set measuring mode, press the button (F) another two times.
In order to perform a measurement, switch on the thermometer as described above and
point the measuring sensor (a) to the surface to be measured at a distance of 3 – 5
cm. Then press the measuring button (D) once. You hear a short beep and can read the
measuring result from the display (b).