Sirona Dental Systems GmbH
Operating Instructions SIROTorque L
Bur Speed
Motor Speed
Motor Speed
Gear Ratio
A
Bur Speed
Motor Speed
Motor Speed
Gear Ratio
B
Bur Speed
Motor Speed
Motor Speed
Gear Ratio
14
C
Bur Speed
Motor Speed
Motor Speed
Gear Ratio
D
Bur Speed
Motor Speed
Motor Speed
Gear Ratio
61 74 499 D 3489
D 3489.201.02.02.02
+
+
SIROTorque L
1
2
Gear Ratio
3
Endo
21
+
SIROTorque L
1
2
Gear Ratio
3
Endo
20
+
SIROTorque L
1
2
Gear Ratio
3
Endo
+
SIROTorque L
1
2
Gear Ratio
3
Endo
21
+
SIROTorque L
1
2
Gear Ratio
3
Endo
23
Instrument light OFF / ON
With key (21) you can switch the instrument light OFF
and ON.
Instrument light ON is signaled by the lamp symbol on
the display.
Changeover with key (21) is also possible while the
motor is running.
After the unit is switched OFF and back ON again, the
last lamp status set remains saved.
i
NOTE
After the foot switch is released, the instrument light
illuminates for approx. 1 second.
Programming handpiece settings
The gear ratio, motor speed and instrument light status
(ON/OFF) can all be reprogrammed together. If you
want to overwrite the factory-programmed values and
settings in the four program keys, proceed as follows:
A
– Select the gear ratio of the required straight/
contra-angle handpiece using the
+/– keys (20). This value then appears on the display.
B
– Select the desired maximum motor speed with the
+/– keys (14). This value then appears on the display,
and also determines the value of the burr speed
C
– Finally, use key (21) to determine whether or not the
instrument light should be switched on.
D
– To program the new values, press the respective
save key (1, 2, 3 or Endo) (23) (in the example shown
here, it would be advisable to press green program key
3 due to the preselected reduction ratio) > 3 seconds.
The new setting is confirmed by a brief acoustic signal.
The newly programmed values can be recalled
whenever required by briefly pressing the corresponding
save key (23).
5 Operation
17