6 Savable settings

6.1 "Limit value adjustment" menu

6
Savable settings
6.1
"Limit value adjustment" menu
Bur Speed
Motor Speed
Motor Speed
Gear Ratio
21
Bur Speed
Motor Speed
Motor Speed
Gear Ratio
20
14
Bur Speed
Motor Speed
Motor Speed
Gear Ratio
Bur Speed
Motor Speed
Motor Speed
Gear Ratio
18
+
SIROTorque L
1
2
Gear Ratio
3
Endo
+
SIROTorque L
1
2
Gear Ratio
3
Endo
+
SIROTorque L
1
2
Gear Ratio
3
Endo
+
SIROTorque L
1
2
Gear Ratio
3
Endo
A
With an incoming air pressure of at least 43 psi
(3 bar) you can work with the factory settings without
having to adjust the limit values.
The limit value adjustments described below only show
the different setting options.
Entering the "limit value adjustment" menu
Briefly press the instrument light key (21) and save
key 3 simultaneously and hold them for at least 3 s.
The parameter
display. At the same time, the value programmed for
this parameter,
You are now in the "limit value adjustment" menu.
Parameter selection is performed using the +/– keys for
the gear ratio (20).
The name of the relevant parameter is shown on the
"Bur Speed" display field.
The adjustment of the value of the selected parameter
within the corresponding limits is performed via the +/–
keys for the motor speed (14). The value of the
parameter is shown in the "Gear Ratio" display field.
As soon as this value is changed, the name of the
parameter starts flashing on the "Bur Speed" display.
Order of parameters with factory-programmed values:
1 Lamp voltage (light intensity)
2 Max. current consumption (torque)
3 Air pressure at which the motor
starts running (lower switching threshold)
4 Air pressure for maximum speed
(upper pressure switching threshold)
5 Reset to factory settings
6 Pressure sensor calibration 1 bar
(no setting required)
7 Pressure sensor calibration 3 bar
(no setting required)
Saving the settings and quitting the menu
A –
Briefly press save key 1 to save the new
parameter. When it has been successfully saved, the
parameter stops flashing.
B –
Briefly press the Endo save key to quit the "limit
value adjustment" programming mode for the system
parameters.
B
Sirona Dental Systems GmbH
Operating Instructions SIROTorque L
then appears on the "Bur Speed"
2.9
, appears on the gear ratio display.
61 74 499 D 3489
D 3489.201.02.02.02
+
BULB 2.9
CU
65
P_LO 1.0
P_HI 3.0
FI_S
0.0
P_P1 0.0
P_P3 0.0