SPECIFICATION
Power Supply:
DET5/4R:
DET5/4D:
Fuses:
DET5/4D:
DET5/4R:
Safety:
E.M.C:
32
Internal rechargeable sealed lead acid cells
12 V, 0,8 Ah capacity. Battery voltage range over which basic
accuracy is maintained, 10,0 V to 13,5 V.
Battery life; 600 x 15 s tests
Battery charging time, 10 hours max. (from completely exhausted).
Charging supply required, (user selectable) 200 V to 255 V a.c. or
100 V to 130 V a.c. 50 Hz/60 Hz.
6 x 1,5 V alkaline battery cells IEC LR6 type. Battery voltage range
over which basic accuracy is maintained, 6 V to 10 V.
Battery life; 600 x 15 s tests.
Internal 100 mA ceramic HBC 20 mm x 5 mm IEC 127/1
(for current source protection)
Internal 100 mA ceramic HBC 20 mm x 5 mm IEC 127/1
(for Potential circuit protection)
Internal 100 mA ceramic HBC 20 mm x 5 mm IEC 127/1
(for current source protection)
Internal 100 mA ceramic HBC 20 mm x 5 mm IEC 127/1
(for Potential circuit protection)
100 mA ceramic HBC 20 mm x 5 mm IEC 127/1 for 240 V a.c.supply,
or 200 mA ceramic HBC 20 mm x 5 mm IEC127/1 for 120 V a.c
supply (for circuit protection during battery charging).
Mains power cord fused plug: 3 Amp fuse to BS 1362
The instrument meets the requirements for safety to IEC 1010-1
1995), EN61010-1 (1995).
In accordance with IEC 61326 including amendment No.1