Out-of-range results
If you continue to have test results fall outside the range printed on the
test strip vial, the meter and strips may not be working properly. Do NOT
test your blood. Go to www.foracare.com or call customer service for
help.
NOTE:
The control solution range printed on the test strip vial is for control solution
use only. It is not a recommended range for your blood glucose level.
See the MAINTENANCE section for important information about your control
solutions.
4. Read and compare the results
After the meter counts to 0, the result of
the control solution test will appear.
Compare the result with the range
printed on the test strip vial. The result
should fall within this range. If not,
please read the instructions again and
repeat the control solution test.
13