Coltene BioSonic US100R Guía De Uso página 13

Ocultar thumbs Ver también para BioSonic US100R:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 36
Owner's Guide
5.4 EN/IEC 60601 - 1-2 :2007 Sub-clause 5.2.2.2 Table 6
Recommended separation distances between portable and mobile RF communications equipment and the Biosonic
The Biosonic
®
US100R unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Biosonic
between the portable and mobile RF communications equipment (transmitters) and the Biosonic
according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter in watts (W)
0.01
0.1
1.0
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
®
US100R unit can help prevent electromagnetic interference by maintaining a minimum distance
Separation distance according to frequency of transmitter in meters (m)
150 kHz to 80 MHz
d = 1.2√P
.12
.38
1.2
3.8
12
US100R unit as recommended below,
®
80 MHz to 800 MHz
d = 1.2√P
.12
.38
1.2
3.8
12
13
US100R unit
®
800 MHz to 2.5 GHz
d = 2.3√P
.23
0,73
2.3
7.3
2.3

Publicidad

Tabla de contenido
loading

Tabla de contenido