Electromagnetic emissions
The medical device is intended for use in an
electromagnetic environment as specified below.
The user must ensure that the medical device is
used in such an environment.
Emissions
Radio frequency emissions (CISPR 11)
Harmonic emissions (IEC 61000-3-2)
Voltage fluctuations/flicker emissions
(IEC 61000-3-3)
30
Compliance according to
Group 1
Class A
Not applicable
Not applicable
Electromagnetic environment
The medical device uses radio frequency energy only for its in-
ternal function. Therefore, its radio frequency emissions are very
low and are not likely to cause any interference in nearby elec-
tronic equipment.
The medical device is suitable for use in all establishments other
than domestic establishments and those directly connected
(without transformer) to the public low-voltage power supply net-
work that supplies buildings used for domestic purposes.