3 Verify PowerPark is charging by confirming that the connection indicator on the stand base glows green.
To undock the stand
1 Make sure the AC power cord is disconnected from the system stand.
2 Wheel the stand away from the dock. The connection indicator turns off.

Troubleshooting

Table 1: Symptoms and solutions
Symptom
PowerPark does not charge the battery
Stand module does not connect to the
dock
To turn the dock circuit breaker on or off
The circuit breaker on the PowerPark dock is typically set to the ON position but trips if there is a power
event (for example, a power surge).
The circuit breaker is a means of disconnection from mains power.
1 Undock the stand.
4
Solution
Make sure that the electrical outlet has power.
Make sure that the dock circuit breaker is turned on.
Reset the stand-module circuit breaker if tripped.
Make sure that the stand module is fully connected to
the dock.
Make sure that the floor is clean, flat, and rigid. Avoid thick
carpeting.