To reduce the risks associated with chemical exposure due to pressure barrier failure, connector failure, purifi cation
media or capsule failure from use of disinfectants within the product: Do not pressurize the capsule more than 5 psi during
introduction of NaOH solution and static soak during sanitization.
3. When fl uid emerges from the vent, close the vent while simultaneously opening the outlet valve slowly.
4. When fl uid emerges from the capsule outlet, stop the fl ow and close the capsule outlet valve.
5. Allow the capsule to incubate with the 1 M NaOH solution. Incubation time should not exceed 1 hour.
6. After the incubation is complete, drain the NaOH solution from the capsule.
Important Notice: Do not blow down the 3M™ Emphaze™ AEX Hybrid Purifi er capsules after NaOH sanitization procedure.
4.B. Required Preconditioning Flush Procedure for the Pre-use NaOH Sanitized Capsules
To reduce the risk of Injury from possible end user exposure associated with degradation products released
from the purifi cation media; and
To reduce the risk of product loss associated with a) chemical release from purifi cation media and b)
nullifi cation of Supplier's USP<88> Class VI compliance of fi lter capsule components.
• The capsule must be subjected to a preconditioning fl ush prior to use.
Perform Required Preconditioning Flush procedure as described in Section 2.A.
IMPORTANT NOTICE: Do not blow down with a gas the 3M Emphaze AEX Hybrid Purifi er capsules after preconditioning
fl ush procedure.
Capsule is ready for immediate use.
WARNING
WARNING