Mechanical alkaline cleaning and thermal disinfecting
Machine type: Single-chamber washer/disinfector without ultrasound
Stage
Step
I
Prerinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfecting
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized)
5.8 Inspection, maintenance and checks
Note
Aesculap recommends lubricating the flap rods, flaps,
sliding surfaces and other moving parts, e.g. trigger,
cover lids etc., with STERILIT® Power Systems oil spray
GB600/STERILIT® Power Systems drip lubricator GA059
prior to sterilization! Remove excess oil with a lint-free
wipe.
Allow the product to cool down to room
temperature.
Inspect the product after each cleaning and
disinfecting cycle to be sure it is clean, functional
and undamaged.
After cleaning/disinfecting, inspect all surfaces and
difficult to access spots on the product for visible
dirt or debris.
Check the product for any damage, atypical
running noise, overheating or excessive vibration.
Set aside the product if it is damaged.
T
t
[°C/°F]
[min]
<25/77
3
55/131
10
>10/50
1
90/194
5
-
-
5.9 Packaging
Follow the instructions for use of the respective
packaging and holders (e.g. instructions for use
TA009721 for the Aesculap Eccos® storage system).
Insert the product in its proper position in the
Eccos® holder, or put it on a tray in such a way that
the product is protected against damage. Make
certain that all cutting edges are protected.
Package trays appropriately for the sterilization
process (e.g. in Aesculap sterile containers).
Ensure that the packaging provides sufficient
protection against recontamination of the product
(DIN EN ISO 11607).
Water
Chemical/Note
quality
D–W
-
FD–W
BBRAUN HELIMATIC CLEANER
alkaline with tensides,
application solution 0.5 %
FD–W
-
FD–W
-
-
according to disinfecting
program