Wipe Disinfection For Electrical Devices Without Sterilization; Manual Cleaning/Disinfecting - Braun Aesculap PV470 Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 64
Aesculap
Full HD CMOS camera system PV470
6.5

Wipe disinfection for electrical devices without sterilization

Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (1 min minimum).
6.6

Manual cleaning/disinfecting

Prior to manual disinfecting, allow water to drip off for a sufficient
length of time to prevent dilution of the disinfecting solution.
After manual cleaning/disinfection, check visible surfaces visually for
residues.
Repeat the cleaning/disinfection process if necessary.
Manual cleaning with immersion disinfection
Phase
Step
I
Disinfecting cleaning
II
Intermediate rinse
III
Disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
RT:
Room temperature
16
®
T
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
T
t
Conc.
[°C/°F]
[min]
[%]
34–45/
5
0,8
95–113
RT (cold) 3x1
-
20–25/
12
0,55
68–77
RT (cold) 3x2
-
RT
-
-
Water quality
Chemical
-
Meliseptol HBV wipes 50 % Propan-1-ol
Water
Chemical
quality
D–W
Enzymatic Cleaner Cidezyme/Enzol
D–W
-
D–W
Cidex OPA (process solution 0.55 % ortho-phtalaldehyde)
FD–W
-
sterile
-
-