Error Messages
MESSAGE
WHAT IT MEANS
Appears when the
batteries cannot provide
enough power for a test.
Appears when a used
test strip is inserted.
Problem in operation.
You may have removed
the strip after applying
blood.
Appears when ambient
temperature is below
system operation range.
Appears when ambient
temperature is above
system operation range.
WHAT TO DO
Replace the batteries
immediately.
Repeat with a new test strip.
Repeat the test with a new
test strip.
If the meter still does not
work, please contact the cus-
tomer service for assistance.
Review the instructions and
repeat test with a new test
strip. If the problem persists,
please contact the local cus-
tomer service for help.
System operation range
is 50°F to 104°F (10°C to
40°C). Repeat the test after
the meter and test strip are
in the above temperature
range.