Problem – – Sample Draw Volume
Problem
Reproducibility - -
Reproducibility
Symptom:
Peak area increases after first injection from the same vial (first injection –
low, latter injections OK)
Possible cause:
FIRST Time.
"Vacuum formation"
injection needle.
Vacuum draws some sample back out of the needle
Situation can be aggravated by over filling the vial
Situation can be aggravated by over filling the vial
o
o
(Never fill the vial all the way to the top)
Test:
Remove cap and septum* from the vial.Perform multiple injections. Measure
peak area to determine if the septum/cap is the
•*
Some auto samplers may have to have the vial sensor disabled
•Vial A
•Inj. 1
•Area = 200,000
Sample Draw Volume
Venting
Venting
Inadequate venting upon needle piercing the septum/cap for the
caused by the septum/cap sealing around the
cause.
•Vial A
•Inj.2
•Area = 250,000
©2007 Waters Corporation
114