Specifications
1-1 General Specifications
Jaw opening : 0.5" (12.7mm)
Accepts up to 1/0 copper cable.
LCD display digits : 6000 digit
large scale LCD readout.
Measuring rate : 5 times / sec.
Overrange display : "OL" is dis-
played for "Ω" function, shows the
real value for "A" and "V" function.
Automatic power off time :
Approximately 30 minutes after
power on.
Low battery indicator : is dis-
played. Replace the battery when
the indicator appears in the
display.
Power requirement : 9V battery.
Battery type and life :
ALKALINE PP3, 9V 250 hours.
1-2 Environmental Conditions
Indoor Use.
Calibration :
One year calibration cycle.
Operating temperature :
0°C ~ 30°C (≦80% RH)
30°C ~ 40°C (≦75% RH)
40°C ~ 50°C (≦45%RH)
Storage temperature :
-20 to +60°C, 0 to 80% RH (batteries
not fitted).
Operating altitude : 2000m (6562 ft)
Overvoltage category :
IEC 61010-1 CAT.Ⅳ600V
CAT.Ⅲ 1000V
IEC 61010-1 600V CAT.Ⅳ.
CAT.Ⅳ equipment of OVERVOLT-
AGE CATEGORY Ⅳ is for use at
the origin of the installations. Note
examples include electricity meters
and primary over-current protection
equipment.
Conductor Size : 27mm diameter.
Pollution degree : 2
EMC : EN 61326-1
Shock vibration :
Sinusoidal vibration per MIL-T-
28800E (5 ~ 55 Hz, 3g maximum).
Drop Protection : 4 foot drop to
hardwood on concrete floor.
8