Indication; Technical Description - MIETHKE proGAV 2.0 Instrucciones De Manejo

Ocultar thumbs Ver también para proGAV 2.0:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 41
GB | INSTRUCTIONS FOR USE
Fig. 1: Schematic cross section of the proGAV 2.0
1. Adjustable differential pressure unit
(adjustable DP-unit)
2. Gravitational unit
3. sapphire ball
4. torsion spring

INDICATION

proGAV 2.0 is used for cerebrospinal fluid
(CSF) drainage in the treatment of hydro-
cephalus.

TECHNICAL DESCRIPTION

The proGAV 2.0 is a valve made from titanium.
It consists of an adjustable differential pres-
sure unit and a gravitational unit.
14
5. rotor
6. tantalum ball
7. sapphire ball
The adjustable differential pressure unit con-
sists of a stable titanium housing in which
a tried and tested ball cone valve (1) is
integrated in the proximal section. A torsion
spring (4) determines the opening pressure
of this unit. The pretension of the spring, and
therefore the valve opening pressure, can be
adjusted through the skin using a revolving
rotor on a bearing (5).
proGAV 2.0

Publicidad

Tabla de contenido
loading

Tabla de contenido