Maintenance

Cleaning EUI Connectors

Regular cleaning of EUI connectors will help maintain optimum performance. There
is no need to disassemble the unit.
If any damage occurs to internal connectors, the module casing will
have to be opened and a new calibration will be required.
To clean EUI connectors:
1. Remove the EUI from the instrument to expose the connector baseplate and
ferrule.
2. Moisten a 2.5 mm cleaning tip with one drop of isopropyl alcohol (alcohol may
leave traces if used abundantly).
3. Slowly insert the cleaning tip into the EUI adapter until it comes out on the other
side (a slow clockwise rotating movement may help).
4
4. Gently turn the cleaning tip one full turn, then continue to turn as you withdraw
it.
Push
3
I
MPORTANT
Turn
5
Pull
14