0.1 L/min. If you are not using the supplied calibration device with pressure reducer the maximum
allowable inlet pressure must be no more than 2 bar absolute.
Note: The manufacturer recommends to keep the flow chamber at atmospheric pressure. Adjust the sample
flow before it goes into the flow chamber.
6. Configure the calibration as described in
7. Start the calibration as described in
(high range sensors only)
preferred calibration method.
7.4.3 Calibration configuration
Note: This option can also be invoked by pressing the Modify button in either the Zero calibration or High level
adjustment calibration screens.
1. From the Main menu, select Calibration followed by Gas sensor and then Configuration.
Option
Auto-calibration
Manual-calibration
Hold during
calibration or
verification
Zero calibration
bottle
Stop parameters
7.4.3.1 Configure manual calibration
1. Set the parameters for a manual sensor calibration:
Option
Calibration mode Select Zero calibration or High level adjustment. If using a high range sensor you also
Cal. sample
Medium
Gas unit type
Gas unit
Zero calibration
on page 22 or
Description
Not available for this sensor.
When Auto-End is enabled a manual calibration will complete automatically when the
parameters defined in Stop parameters are reached. Press on Configure to set the
manual calibration parameters. If the calibration fails, the previous calibration
parameters remain unchanged and a warning message displayed.
If checked this keeps the last measured value and stops updating the outputs during
the calibration or verification process. This avoids sending invalid information to any
connected device. At the end of a calibration, this hold remains on for a further
10 minutes to allow the system to stabilize.
Make sure this is disabled by unchecking the box as it is not relevant for this sensor.
If this button is pressed you can view or change the existing values or restore the
default values. It is highly recommended to leave these parameters at their default
values. These values apply to manual calibrations with the Auto-End parameter
enabled.
Description
have the option of 100% humid air calibration.
Note: If zero calibration or 100% humid air calibration is selected, no other parameters
are required. The following are only required for high level adjustment.
Set to In line sample, Gas bottle or Factory parameters. If factory parameters is
selected, the Ksv value is displayed but can be changed.
These additional parameters are required if in line sample or gas bottle has been
selected as the calibration sample:
This is automatically set to Liquid if in line sample has been selected as the calibration
sample, or Gas if gas bottle has been selected.
Either Partial or Dissolved are available for an in line sample. If gas bottle was selected
this is set to Fraction.
The list of available units depends on unit type selected above.
Calibration configuration
on page 22,
High level adjustment
on page 21.
100% humid air calibration
on page 22 depending on the
English 21