Specification
Repeatability
Reproducibility
Accuracy
Limit of detection (LOD)
Response time (90%)
Display resolution
Calibration
Calibration sample
Ambient temperature
Humidity
Power supply
Wall and pipe mount
instrument
(H x D x W)
Panel mount instrument
(housing) (H x D x W)
K1200 28 mm sensor
(L x W)
Calibration device
Section 2 Expanded manual version
For additional information, refer to the expanded version of this manual, which is available on the
manufacturer's website.
Section 3 General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.

3.1 Safety information

The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
4 English
Details
Low range sensors: ± 0.4 ppb or 1%, whichever is the greater
High range sensors: ± 0.015 ppm or ± 2% whichever is the greater
Low range sensors: ± 0.8 ppb or 2%, whichever is the greater
High range sensors: ± 0.02 ppm or ± 3% whichever is the greater
Low range sensors: ± 0.8 ppb or 2%, whichever is the greater
High range sensors: ± 0.02 ppm or ± 3% whichever is the greater
Low range sensors: 0.6 ppb
High range sensors: 0.015 ppm
Low range sensors: < 10 seconds in gas phase; < 30 seconds in liquid phase
High range sensors: < 10 seconds in gas phase; < 50 seconds in liquid phase
0.1 ppb
Low range sensors: Single point calibration (zero)
High range sensors: Two at cap replacement (zero and air), one during use (air)
Low range sensors: Standard 99.999% N
gas
High range sensors: Standard 99.999% N
gas, air
-5 to 50°C (23 to 122°F)
0 to 95% non-condensing relative humidity
Universal 85-264 VAC @ 50/60 Hz, 25 VA
10-36 VDC, 25 W
236.5 x 160 x 250 mm; Weight 4.25 kg
9.31 x 6.30 x 9.84 in.; Weight 8.82 lbs
156 (123) x 250 x 220 (214) mm; Weight 3.35 kg
6.14 (4.84) x 9.84 x 8.86 (8.43) in.; Weight 6.62 lbs
143.50 x 49 mm; Weight 0.74 kg
5.65 x 1.93 in.; Weight 1.63 lbs
Weight 0.7 kg (1.54 lbs)
N O T I C E
(quality 50) or equivalent oxygen free
2
(quality 30) or equivalent oxygen free
2