Saving, Retrieving And Deleting Results; Error Messages / Troubleshooting; Cleaning And Care; Technical Details - Beurer medical BC 32 Instrucciones Para El Uso

Ocultar thumbs Ver también para BC 32:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 25
If the values for systolic and diastolic pressure are in two dif­
ferent WHO ranges (e.g. systolic in the high­normal range and
diastolic pressure in the normal range) the graphic WHO clas­
sification on the unit indicates the higher range (high­normal in
the example described).

6. Saving, retrieving and deleting results

• The results of each successful measurement are stored
together with date and time. With more than 60 items of
measured data, the earliest items of data measured are lost.
• Select the desired user memory by pressing the MEM button
and then the + button. Press the MEM button again to display
the average of all saved measurements in the user memory.
Press the MEM button again to display the average day
measurement for the last 7 days. (day: 5.00 a.m. – 9.00 a.m.,
display A). Press the MEM button again to display the aver-
age night measurement for the last 7 days
(night: 5.00 p.m. – 9.00 p.m., display P). If you continue
pressing the MEM button, the latest individual results are
displayed with date and time.
• To clear the memory, press the MEM button, then the display
shows no. 1. Press the + button to select the user memory
and confirm by pressing MEM again. Press and hold the „+"
and „ " keys simultaneously for 5 seconds (the display will
show CLA).
• If you wish to change the user memory, see "Select memo-
ry".

7. error messages / troubleshooting

In case of faults, the E_ message appears in the display.
Error messages can occur when
1. inflation pressure is higher than 300 mmHg E2.
2. blood pressure is unusually high or low E3.
3. you move or talk during the measurement (in addition to E_,
the cardiac arrhythmia icon „
4. the cuff tube is not inserted properly E1.
5. the cuff is inflated for longer than 25 seconds E1.
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
not move or talk. Re-insert the batteries if necessary, or else
replace them.

8. Cleaning and Care

• Clean your blood pressure monitor carefully using only a
slightly moistened cloth.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise liquid
can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.

9. Technical details

Model no.
Measurement method: Oscillating, non­invasive blood
Measurement range
Precision of
pressure reading
15
" appears in the display E3),
BC 32
pressure measurement on the wrist
systolic 40 – 280 mmHg,
diastolic 40 – 280 mmHg,
pulse 40 –199 beats/minute
Systolic ± 3 mmHg /
diastolic ± 3 mmHg
pulse ± 5% of displayed value

Publicidad

Tabla de contenido
loading

Tabla de contenido