Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 4)
Electromagnetic immunity
The electric breastpump Freestyle is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breastpump Freestyle should assure
that it is used in such an environment.
IEC 60601
Immunity test
test level
3 Vrms
Conducted RF
150 kHz to
IEC 61000-4-6
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2,5 GHz
Compliance
Electromagnetic
level
environment – guidance
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the electric
breastpump Freestyle, including ca-
bles, than the recommended
separation distance calculated from
the equation applicable to the fre-
quency of the transmitter.
Recommended separation distance
d = 0,35√P
d = 0,35√P
d = 0,7√P
10 Vrms
where P is the maximum output
10 V/m
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF trans-
mitters, as determined by an electro-
magnetic site survey,
than the compliance level in each fre-
quency range.
Interference may occur in the vicinity
of equipment marked with the follow-
ing symbol:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
a
should be less
b
147