isokinetic probe supplies the same vertical (or horizontal) flow speed of
the air in order to collect representative samples of the clean room
laminar flow for the instrument. Refer to
sampling with and without the isokinetic probe.
Figure 3 Isokinetic probe function
1 No probe in non-laminar air flow
2 To particle counter

Product components

Make sure that all components have been received. Refer to
any items are missing or damaged, contact the manufacturer or a sales
representative immediately.
Figure 3
for a comparison of
3 No probe in laminar air flow—
particles are missed
4 Isokinetic probe in laminar air flow
—most accurate
Figure
4. If
Figure 4 Instrument components
1 MET ONE 6000P series particle
counter
2 Sample (isokinetic) probe with tube
3 Zero count filter
1
4 115 VAC power cord
1
5 230 VAC power cord
1
AC units only
2
RS485 units only
3
Ethernet units with DC input or Ethernet units with analog output only
4
Only one service port cable is supplied per order.
Installation
Installation guidelines
N O T I C E
This instrument is not waterproof or resistant to vaporous hydrogen peroxide
(VHP).
6 8-pin connector for DC input power
2
and serial communications
7 7-pin connector for DC input power
3
and analog output
8 Service port cable (8-pin DIN to 9-
4
pin serial connector)
9 Setup utility CD
English 7