Technical Data - Etac Molifit Air Tift Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 58
Molift Air Tilt / www.etac.com

Technical data

Safe Working Load (SWL)
272 kg (600 lbs)
Weight of unit (excluding hand control and cables):
7,3 kg (16 lbs)
Actuator speed
12,6-10,4mm/second (0,49-0,40 inches/second)
Protection class:
Hand control IP24
Actuator IPX4
Operating forces button
Buttons on handset: 3.4 N
Material:
Steel, Plastic
Expected Lifetime:
Expected lifetime of 30 000 cycles SWL or 10 years.
Dimensions:
L x W x H (Length, Width)
66 x 58 x 50-79 cm
(26 x 22,8 x 19,7-33,5 inches)
Tilting Range:
75 degrees
47cm
58cm
Electronics
Transport and Storage
The hoist can be stored and transported under temperatures
between -25 - 70 °C.
Operating
Designed for use at standard room temperatures (+5 to +40°C).
Air Pressure: 70 - 106 kPa
Relative Humidity: 15 - 93 %
After storage or transport at other temperatures leave it in a
room with a suitable temperature until it reaches a safe operating
temperature.
Medical electrical equipment requires special
precautions regarding electromagnetic
compatibility (EMC). Portable or mobile radio
communication equipment may affect the
medical electrical equipment, and should
be kept minimum 25 cm (10 inches) from the
hoists electronics.
Working pause ratio/Duty Cycle.
Duty Cycle 10%. (Intermittence according to standard ISO-EN
10535)
58cm
75°
3

Publicidad

Tabla de contenido
loading

Tabla de contenido