TROUBLESHOOTING

If any abnormality arises during use, please check the following points:
Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no
IB-WGNBPA-750--03.indd 35
SYMPTOMS
Unit does not turn
on when the START/
STOP button is
pushed.
EE measurement error
symbol shown on dis-
play or the blood pres-
sure value is displayed
excessively low
(or high).
E1 error symbol shown
on display
E2 error symbol shown
on display
E3 error symbol shown
on display
EP error symbol shown
on display
circumstance should you disassemble or attempt to repair the unit by yourself.
Contact information for HoMedics Consumer Relations Department can be
found on the warranty page.
POSSIBLE
CAUSES
Batteries have
run down.
Battery polarities
have been positioned
incorrectly.
Cuff has been placed
incorrectly.
Did you talk or move
during measurement?
Shaking of the arm with
the cuff on.
Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.
Inflation pressure
exceeding 300 mmHg.
Error determining
measurement data.
System error.
35
CORRECTION
Replace them with
four new AA
alkaline batteries.
Re-insert the batteries
in the correct positions.
Wrap the cuff properly
so that it is positioned
correctly. Measure
again.
Keep arm steady
during measurement.
Measure again.
Check cuff connection.
Measure again.
Switch the unit off, then
measure again.
Measure again.
Measure again.