ALARM PARAMETERS

Setting temperature alarm delay

This function delays the activation of the temperature alarm for the preset amount of time.
Settable range: 0 – 15 minutes
Factory setting: 15 minutes
1. On the home screen, press the menu key (MENU/HOME).
▶ The left side of the display changes to the Menu screen.
2. On the Menu screen, move the cursor to Alarms & Controls [Fig. 1]
using the up/down keys (△▽), and press the enter key (ENTER).
▶ The display changes to the Alarms & Controls screen.
3. When the cursor in the Alarms & Controls screen is on Temp/Gas
Alarm Set [Fig. 2], press the enter key (ENTER).
▶ The display changes to the Temp/Gas Alarm Set screen.
4. On the Temp/Gas Alarm Set screen, move the cursor to Temp Alarm
Delay [Fig. 3] using the up/down keys (△▽), and press the enter key
(ENTER).
▶ The right side of the display changes to the Temp Alarm Delay screen,
and the current setting value (15) is displayed [Fig. 4].
5. Move the cursor using the left/right keys (◁▷) and select the digit you
want to change, and then use the up/down keys (△▽) to change the
number.
6. Press the enter key (ENTER).
▶ The value is confirmed and the display returns to the Temp/Gas Alarm
Set screen.
*If the entered setting value is out of the settable range, the following
message is displayed on the right side of the display. Press any key on
the control panel to change to the Temp Alarm Delay screen [Fig. 4].
7. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
Note: When the incubator is recovered from the alarm state within the set alarm delay time, the buzzer
doesn't sound after the elapse of the set alarm delay time.
56
Temp Alarm Delay
Invalid Value
Press Any Key
0min – 15min
<Right side of the display>
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Left side of the display>
Alarms & Controls
Temp/Gas Alarm Set
Other Alarm Set
UV Setting
Gas Supply Setting
Date & Time
Keypad Click Set
<Left side of the display>
Temp/Gas Alarm Set
Temp Alarm
Temp Alarm Delay
CO
Alarm
2
CO
Alarm Delay
2
<Left side of the display>
Temp Alarm Delay
15
min
0min – 15min
<Right side of the display>
[Fig. 1]
[Fig. 2]
[Fig. 3]
[Fig. 4]