UV LAMP (OPTION)

Setting UV lamp ON period

To change the UV lamp ON period, follow the procedure below.
 Settable range: 0 – 30 minutes
 Factory setting: 10 minutes
Notes:
・The recommended UV lighting time (UV Lighting Time) is 10 minutes. Less than 10 minutes may reduce
the bactericidal effect.
・When set to 0 minutes, the UV lamp does not light up.
1. On the home screen, press the menu key (MENU/HOME).
▶ The left side of the display changes to the Menu screen.
2. On the Menu screen, move the cursor to Alarms & Controls [Fig. 1]
using the up/down keys (△▽), and press the enter key (ENTER).
▶ The display changes to the Alarms & Controls screen.
3. On the Alarms & Controls screen, move the cursor to UV Setting [Fig.
2] using the up/down keys (△▽), and press the enter key (ENTER).
▶ The display changes to the UV Setting screen.
4. With the cursor in the UV Setting screen on UV Lighting Time [Fig. 3],
press the enter key (ENTER).
▶ The right side of the display changes to the UV Lighting Time screen,
and the current setting value (10) is displayed [Fig. 4].
5. Move the cursor using the left/right keys (◁▷) and select the digit you
want to change, and then use the up/down keys (△▽) to change the
number.
6. Press the enter key (ENTER).
▶ The value is confirmed and the display returns to the UV Setting
screen.
*If the entered setting value is out of the settable range, the following
message is displayed on the right side of the display. Press any key on
the control panel to return to the UV Lighting Time screen [Fig. 4].
7. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
UV Lighting Time
Invalid Value
Press Any Key
0min – 30min
<Right side of the display>
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Left side of the display>
Alarms & Controls
Temp/Gas Alarm Set
Other Alarm Set
UV Setting
Gas Supply Setting
Date & Time
Keypad Click Set
<Left side of the display>
UV Setting
UV Lighting Time
UV Life Counter
Auto-Extended Time
UV 24h Mode Start
<Left side of the display>
UV Lighting Time
10
min
0min – 30min
<Right side of the display>
[Fig. 1]
[Fig. 2]
[Fig. 3]
[Fig. 4]