H
O
DECONTAMINATION (OPTION)
2
2
*If an error occurs during the
system check, the error screen is
displayed [Fig. 9].
Check the error, and then stop the
product and contact our sales
representative or agent, unless the
cause of the error is clear and
recovery is expected.
14. When H
O
decontamination
2
2
is complete, the Decon Finished
message is displayed [Fig. 10].
Open the outer door and inner door,
remove the power connector of the
H
O
generator from the connector
2
2
of the main unit, and then remove
the H
O
generator. The right side
2
2
of the display switches to the
screen shown in Fig. 11.
Note:
Always
wear
glasses and rubber gloves when
performing this procedure.
15. Press the Enter key (ENTER).
▶ The Home screen is displayed.
16. Dilute the remaining H
large amount of water for disposal, and then rinse and wash the
H
O
generator with distilled water. Store the H
2
2
outside of the incubator in a clean environment.
Note: Secure the waterproof connector cap firmly to the connector
for the H
O
generator to protect the connector part from getting wet
2
2
[Fig. 12].
Do not clean inside and outside of the H
for disinfection.
17. After the H
O
2
2
in the bottom of the chamber and the lower part of the H
generator. Therefore, wear protective glasses and rubber gloves
and then, wipe clean these parts with a non-woven cloth.
18. Ventilate the chamber sufficiently, and then set the inner
attachments back in their original positions in the chamber.
Notes:
ꞏ After the H
O
2
2
connector cap firmly to the connector of the incubator [Fig. 13].
ꞏ Make sure not to lose the lid of the H
Caution: Wait at least one hour before starting the cultivation until
the culture environment is stable.
76
<Left side of the display>
<Left side of the display>
protective
<Left side of the display>
O
reagent in the H
2
2
2
decontamination work, the H
decontamination work, secure the waterproof
O
2
2
H
O
Decontamination
2
2
Temp.
Door : Closed
37.0
UV : OFF
05/10
◀
▶
H
O
Decontamination
2
2
Temp.
Door : Closed
48.0
UV : OFF
H
O
Decontamination
2
2
Temp.
Door : Closed
48.0
UV : OFF
O
generator with a
2
2
O
generator
2
2
O
generator with alcohol
2
O
liquid will remain
2
2
O
2
2
generator.
Decon Failed.
Please Remove
H
O
2
2
Power Failure
Err35 :
<Right side of the display>
Decon Finished.
Please Remove
H
O
2
2
<Right side of the display> [Fig. 10]
Decon Finished.
Please Press Enter Key.
<Right side of the display>
Unit.
[Fig. 9]
Unit.
[Fig. 11]
Waterproof
Connector Cap
Fig. 12