H 2 O 2 Decontamination (Option); H 2 O 2 Decontamination - Phcbi MCO-50AIC Instrucciones De Operación

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

H
O
DECONTAMINATION (OPTION)
2
2
When the chamber is contaminated or when cleaning the chamber prior to starting a culture, H
decontamination can be performed.
Prior to the H
O
decontamination work, make sure that MCO-170UVS (UV system set), MCO-50HP (H
2
2
generator), MCO-50HB (H
Notes:
● Before performing decontamination, read the precautions for safe operation on pages 5– 9.
● The door is locked by the electric lock during decontamination. However, if the key included with the
electric lock has been inserted into the key hole and left in the position, the door cannot be locked. When
performing decontamination, be sure to take the key out of the key hole and store and manage the key in a
safe place.
H
O
decontamination
2
2
1. Remove the trays, fan cover, duct, and humidifying pan from the chamber.
2. Then wipe clean the removed inner attachments with alcohol for disinfection.
3. Wipe clean the chamber with alcohol for disinfection.
4. Attach the duct and fan cover.
5. Pour a bottle of MCO-5H2O2 H
generator and then, cover it with the lid.
Note: H
O
reagent cannot be stored after opening.
2
2
Note: Follow the cautions below when using the optional MCO-5H2O2 (H
● Cautions in handling
- Wear protective equipment such as protective glasses and rubber gloves when handling the H
- Do not recklessly use fire in an area where H
- Do not leave H
O
2
- Do not place combustible or inflammable objects near areas where the H
● Cautions in storage
- Store the H
O
reagent in a cool, dark place.
2
2
- Always tighten the cap to the container to prevent contamination by adulterants.
- Make sure the container is free of damage such as corrosion or cracking.
- Store the containers with the mouth facing upwards, and make sure it cannot topple or fall.
● Cautions in disposal
- Dilute the H
O
reagent with a large volume of tap water for disposal.
2
2
6. Connect one end of the waterproof cable (included with the H
and set the positioning fittings of the H
Ensure that the H
decontamination.
7. Remove the waterproof connector cap from the connector on the rear right of the chamber, and then
connect the other end of waterproof cable to the connector (Fig. 2).
Note: Make sure not to lose the waterproof connector cap.
Slits
Duct
Positioning fittings
74
O
decon kit), and MCO-170EL (electric lock) are all installed on your incubator.
2
2
O
reagent (designated by our company) into the MCO-50HP H
2
2
reagent in the container after or during use.
2
O
generator into the slits of the duct (Fig. 1).
2
2
O
generator is properly set. An improper setting may cause inadequate
2
2
Waterproof
H
O
2
2
O
reagent is handled.
2
2
Front
cable
generator
Fig. 1
O
reagent).
2
2
O
reagent is handled.
2
2
O
generator) with the H
2
2
O
2
2
O
2
2
O
2
2
O
reagent.
2
2
O
generator
2
2
Connector
Waterproof
Cable
Fig. 2

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido