Filter assembling
Make sure the filter is assembled with the arrows on the side of the patient.
WARNING: The inside of the medical device must be regularly checked for the presence of liquids
or other visible contamination (secretions). In the presence of liquids or other visible contamination,
immediately replace the medical device due to the risk of an insufficient vacuum flow rate.
These products have been designed, tested and manufactured exclusively for single patient use and
for a period no longer than 24 hours.
Operation with Internal Battery
• Press the switch in position I to turn the device on (the external power pack doesn't need to be connected)
• The fully charged battery life is about 60 minutes with continuous operation.
WARNING: Before using the device, check the battery power status. Before each use proceed with
charging the battery. To maintain the device in good conditions, recharge the battery every 3 months (when
not in use).
Recharging operations:
switching adapter to the electric network for approx. 240 minutes with the main switch to position 0.
TAB. I – INDICATOR LIGHTS DURING OPERATIONS
When an external power supply (regardless of the state of the battery charger) and when the device is work-
ing (after having turned it on), the LED stays in a FIXED GREEN position.
LED Signal
Phase
Flashing Green
During rechanrge
Led
Steady Greed Led
During rechanrge
Steady Red Led
During battery
operation
Flashing Red Led
Device automatically
turns off when the
battery is flat
Steady Orange Led During battery
operation
to be able to charge the internal battery it is necessary to connect the universal
Problem / Cause
Battery recharge running
Recharging cycle
complete
Flat battery
Battery completely flat
Intermediate status
ENGLISH
Solution
Wait
Remove power supply
Start recharging cycle
WARNING: During this signal, you will hear a
long, continuous beep (duration of sound 0.8
sec / sound frequency: every 8.5 sec), which
notifies the user regarding the battery dis-
charge.
When the device is restarted the LED will flash
red: begin the battery recharge cycle immedi-
ately
Guaranteed battery function / Recharge when
the red LED signal comes on.