17
Working Condition
Conservation condition and Transport
The technical specifications may change without notice.
CLEANING OF THE DEVICE
Use a soft dry cloth with not – abrasive and not – solvent detergents. To clean the device external parts always
use a cotton cloth dampened with detergent. Don't use abrasive or solvent detergents. Before carrying out
any cleaning and / or maintenance operation, disconnect the appliance from the power supply, unplugging it
or turning off the switch on the device.
Particular care should be taken to ensure that the internal parts of the equipment do not get in
touch with liquids. Never clean the equipment under water.
During all clearing operations use protection gloves and apron (if need be, also wear a face mask and glass-
es) to avoid getting in contact with contaminating substances (after each utilization cycle of the machine).
ACCESSORIES SUPPLIES
• N°2 complete aspiration jar 2000ml
• Conical fitting
• Tubes set 8 mm x 14 mm
• Hydrophobic and antibacterial filter
• AC/DC adapter
• European power supply cord for AC/DC adapter
The filter is produced with (PTFE) hydrophobic material to prevent fluids entering the pneumatic circuit.
It should be changed immediately if it becomes wet or if there is any sign of contamination or discolouration.
If should also be changed if the unit is used with a patient whose risk of contamination is unknown. Don't use
the suction unit without the protection filter. If the suction unit is used in an emergency or in a patient where
the risk of contamination is not known the filter must be changed after each use.
The filter is not designed to be decontaminated, disassembled and/or sterilised. If the patient's pathology is
known and/or no risk of indirect contamination exists, it is advisable to replace the filter after each work shift
or at any rate on a monthly basis even if the device is not used.
WARNING: Suction tubes for insertion in the human body purchased separately from the machine
should comply with ISO 10993-1 standards on material biocompatibility.
Aspiration jar: the mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and
sterilization. Beyond this limit, the physical-chemical characteristics of the plastic material may show signs of
decay. Therefore, we recommend that you change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the
said tube. Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable
for reuse. The component must be replaced if there are visible signs of decay of the material constituting the
said component.
Conical fitting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked to
the employment of the said component. Therefore, after each cleaning cycle, it is up to the final user to verify
whether the fitting is suitable for reuse. The component must be replaced if there are visible signs of decay in
the material constituting the said component.
ENGLISH
Room Temperature: 5 ÷ 35°C
Room Humidity Percentage: 30 ÷ 75% RH
Atmospheric pressure: 800 ÷ 1060 hPa
Altitude: 0 ÷ 2000m s.l.m.
Room Temperature: - 40°C ÷ 70°C
Room Humidity Percentage: 10 ÷ 100% RH
Atmospheric pressure: 500 ÷ 1060 hPa