Configuration - Dräger Atlan A300 Manual Del Usuario

Ocultar thumbs Ver también para Atlan A300:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 15
|
en

Configuration

Configuration
System setup > System
Vertical tab "Interfaces"
The following information has been updated:
Headline/
Parameter
COM 1
Protocol
Baud rate 1200; 2400; 4800; 9600; 19200; 38400
COM 2
Protocol
Baud rate 1200; 2400; 4800; 9600; 19200; 38400
1) In the case of a data connection using data that does not comply with MEDIBUS,
no data is transmitted to the connected device. Delayed graphs and deviations in
the airway pressure displayed on the screen and in the status display of the mixer
can occur on the anesthesia machine.
Set identical baud rates on all connected devices for each data connection.
Technical data
Operating characteristics
The following information has been updated:
Protection classes
Degree of protection
6
Setting range
MEDIBUS.X; None
MEDIBUS.X; None
IP11 against vertically falling dripping water under
typical conditions of use in accordance with the
intended purpose (with 0° inclination of the device)
IP10 if the device has an inclination of >0° and ≤5°
In accordance with IEC 60529, meets ISO 80601-
2-13
|
Supplement
Description
Configures the COM port.
A baud rate of 19200 or 38400 is
required for transmission of high-speed
data, e.g., for waveforms.
If it cannot be ensured that the
connected device sends MEDIBUS-
1)
compliant data
, set a baud rate of
9600.
Atlan A300, A300 XL, A350, A350 XL SW 1.0n

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Atlan a300 xlAtlan a350Atlan a350 xl

Tabla de contenido