Routine maintenance includes cleaning or replacing the filter as needed. The
Filter is located under the top of the housing in the back. A finger bolt and
plate keep the filter in place. Visually inspect filter for dirt or heavy debris
buildup such as hair and dirt. The Filter can be washed with hand soap and
air-dried. Make certain the filter is dry before replacing it. A brittle or
damaged filter MUST always be replaced.
Periodic testing will verify proper Jay Dermafloat LAL System operation.
Inflation pressures lower than specified can result in poor or reduced patient
support.
NOTE: All tests or service should be performed only by qualified
Technical Personnel.
Maintenance Test Schedule:
The following test should be performed every 6 to 9 months. Keep
written records of each test.
Electrical Leakage Current Test:
To perform an electrical leakage current test, an electrical leakage
analyzer such as those manufactured by "Bio-Tec" or "Cam-Safe"
should be used. Follow the manufacturer's instructions for setting
up and performing the tests required to determine ground resistance
and current leakage.
For safe performance, the unit should not be used if the impedance
or leakage current values exceed the maximum allowable values.
The maximum ground impedance allowable by UL is 0.1 Ohms
with a maximum leakage current of 100 micro amperes (500 micro
amperes for 230V Model).