7. CALIBRATION PROCEDURE

Switch on the balance, and press the CAL switch during the self-test.
When the self-test procedure finished, the display will show as below:
When it shows "CAL" , press
Touch the pan lightly by hand, and several seconds later, once it take stability again, the display shows the
weight value of first calibration mass (1/3 of full capacity). Place the displayed weigh on the pan.
Several seconds later, once it take stability, the display shows the weight value of second calibration mass
(2/3 of full capacity) Place the displayed weigh on the pan.
18
TWB
V3
GWA
V4
GWB
V5
QUA
V6
QUB
V7
UWA
V8
UWB
V9
UNT
V10
CAL SWITCH
CAL
key to start calibration procedure.