4. FUNCTIONS AND OPERATION

The balance can be available for operation from mains at 110~240V (50/60Hz) with an external AC/DC
adapter with 12V output voltage and an optional internal 6V rechargeable battery.

4.1 POWER-UP

When power up the balance, the display will first show the software version for 2 seconds and then per-
form a self display test. Then it will show the counting times for about 1 second. After that it will automati-
cally establish the current weight as a new zero reference point.

4.2 WEIGHING MODE

4.2.1 DISPLAY RANGE

The display can show weight Tare weight to Max+9e (Gross weight).

4.2.2 ZERO-SETTING

Zero-setting range: ±2% of Max
Zero-setting is only available when the load receptor is not in motion (The display is stable).
SEMI-AUTOMATIC ZERO-SETTING
When there is a minor weight displayed (without load on the pan), press ZERO key to return to the
display to zero, then the Zero symbol appears which indicates that the display now is at the zero point.
ZERO-TRACKING
The instrument is equipped with a zero-tracking feature which operates at a speed of ±0.25e/s and only
when the indicator is at gross zero and there is no motion in the weight display.
INITIAL ZERO-SETTING
Initial zero-setting range: ±10% of Max
When the initial zero-setting range is below -10% of Max, the display shows "-------" and the buzzer
sounds. To return the display to normal mode, please adding more weights until the initial zero-setting
range is between ±10% of Max.
When the initial zero-setting range is beyond +10% of Max, the display shows "-------" and the buzzer
sounds. To return the display to normal mode, please take off weights until the initial zero-setting range
is between ±10% of Max.
USER MANUAL BAP
To return to last step in setting
To move the digit to right
To confirm the current setting in setting
mode
EN