OPERATION/ALARM LOG
Note: When the specified operation log data does
not exist, a dialog box is displayed. Press OK, and
select a new period following the procedure 5 and
6.
7.
When the export is complete, a dialog box is
displayed. Press OK.
Note: Even after the export is completed,
operation log data saved in the unit is not deleted.
8.
Remove the USB memory device from the USB port.
Note:
•A log folder is created in the USB memory device, and the exported file is saved in this folder in CSV
format. The name of the exported file is in the format; date (8 digits) - type of data.
(e.g.) When exporting all types of data using All (from Jan. 1st, 2017 to Oct. 1st, 2017):
20170101-20171001_AllCh.csv
20170101-20171001_Door.csv
(e.g.) When exporting Actual Temp. using 1 Day (Jan. 1st, 2017):
20170101_Temp.csv
If the file name is duplicated, A sequential number such as "-1" is added to the end of the file name to be
*
output later.
•At the beginning of the exported file, product name ("MDF-DU302VX") is written. However when the
Unique ID is registered [page 33], product name and Unique ID (8-digit) are written.
(e.g.) When "RoomA001" is set as the Unique ID of MDF-DU302VX:
MDF-DU302VX, RoomA001
9.
Press Top to return to the Top screen.