OPERATION/ALARM LOG
4.
Alarm logs from the last 7 days (including
current day) are displayed.
Note: When the number of alarm log exceeds 6,
pressing the top (▲) or the bottom (▼) log scrolls
the log table and hidden alarm logs can be seen.
•Press Back to return to the Log screen.
•Press Top to return to the Top screen.
5.
Change the number in the Last XX Days input
box to display alarm logs for specified days
(including current day)
Settable range: 1 day~45 days.
Note: Accurate to about 1 minute per month. Refer
to page 44 for the procedure of setting time.
•Press Back to return to the Log screen.
•Press Top to return to the Top screen.
●
From the Alarm screen, alarm log data can be exported in CSV format to the USB memory device.
6.
Insert the USB memorydevice into the USB port.
Note: It is not possible to use a USB memory device with a security function that requires entering a
password.
7.
Press Export.
8.
When the export is completed, a dialog box is
displayed. Press OK. Refer to pages 42 and 43 for
the details about abnormal export or name of
exported file.
9.
Press Top to return to the Top screen.