3.1. General rules

3.1.3 Status messages

Go:
$G
Hold:
$H
Continue:
$C
Ready:
$R
Stop:
$S
Detailed status conditions
Status conditions of the global $G:
$G .Mode.DET .Inac:
.Req .Id1:
.Start:
.Titr:
$G .Mode.MET...
$G .Mode.SET .Inac:
.Req .Id1:
.Start:
.SET1:
.SET2:
.Cond.Ok:
.Cond.Prog:
$G .Mode.KFT...
$G .Mode.STAT .Inac:
.Req .Id1:
.Start:
.Titr:
.Wait:
$G .Mode.DOS...
108
In order to have an efficient control by an external control device, it must also
be possible to query status conditions; they provide information on the status
of the Titrino. The trigger $D initiates output of the status. Status messages
consist of the global status, the detailed status and eventual error messages,
e.g. $S.Mode.SET;E26. The global status informs on the activity of the proc-
ess, while the detailed status conditions show the exact activity within the
process.
The following global status conditions are possible:
The Titrino is executing the last command.
The Titrino has been held ($H, key <meas/hold> or by an error which effects
the hold status)
The Titrino has been restarted actively after hold
The Titrino has executed the last command and is ready
A process has been aborted in an "unnatural manner". e.g. stopped or aborted
because there was an error.
Instrument at the beginning or at the end of a titration.
Instrument in the DET mode, requesting Id1 after titration start.
Instrument in the DET mode, requesting Id2 after titration start.
.Id2:
Instrument in the DET mode, requesting Id3 after titration start.
.Id3:
Instrument in the DET mode, requesting sample size after titration start.
.Smpl:
Instrument in the DET mode, requesting unit of sample size after titration
.Unit:
start.
Instrument in the DET mode, processing the start conditions.
Instrument in the DET mode, titrating.
As DET.
Instrument at the beginning or at the end of a titration.
Instrument in the SET mode, requesting Id1 after start.
Instrument in the SET mode, requesting Id2 after start.
.Id2:
Instrument in the SET mode, requesting Id3 after start.
.Id3:
Instrument in the SET mode, requesting sample size after start.
.Smpl:
Instrument in the SET mode, requesting unit of sample size after start.
.Unit:
Instrument in the SET mode, processing the start conditions.
Instrument in the SET mode, titrating to the first endpoint.
Instrument in the SET mode, titrating to the second endpoint.
Instrument in the SET, conditioning, endpoint reached (after the first
startup from the standby mode).
Instrument in the SET mode, conditioning, endpoint not reached
(Conditioning progressing).
As SET.
Instrument at the beginning or at the end of a determination.
Instrument in the STAT mode, requesting Id1 after start.
Instrument in the STAT mode, requesting Id2 after start.
.Id2:
Instrument in the STAT mode, requesting Id3 after start.
.Id3:
Instrument in the STAT mode, requesting sample size after start.
.Smpl:
Instrument in the STAT mode, requesting unit of sample size after start.
.Unit:
Instrument in the STAT mode, processing the start conditions.
Instrument in the STAT mode, controlling.
Instrument in the STAT mode, waiting because a monitored limit has
been violated.
As STAT.
799 GPT Titrino