3.2. Remote control commands
166
statistics, silo, calib, C-fmla, def, user method, full, short, mplist,
curve, deriv, comb, meas crv, temp crv, adj para, scalc full,
$G sends the selected report to the COM which is set in
&Config.PeriphUnit.RepToComport:
configuration: Configuration report. Is not accessible during a running deter-
mination.
parameters: Parameter report of the current method. During a running deter-
mination only "live"-parameters are accessible.
smpl data: Current sample data.
statistics: Statistics table with the individual results.
silo:
Contents of the silo memory.
calib:
Calibration data of the measuring input in the current method.
C-fmla:
Contents of the <C-fmla> key.
def:
Contents of the <def> key.
user method: Contents of the method memory.
full:
Full result report of the last completed determination.
short:
Short result report of the last completed determination.
mplist:
Measuring point list of the running determination.
curve:
Titration curve of the last determination.
derive:
1st derive of titration curve of the last determination (with DET).
comb:
1st derive combined with the titration curve of the last determina-
tion (with DET).
meas crv: Curve measured value vs. time (with STAT, DOS, DOC and acti-
vated meas monitoring).
temp crv: Curve temperature vs. time (with STAT, DOS, DOC and activated
temperature monitoring).
adj para:
Adjustment parameters.
scalc full: Full report of the silo calculations.
scalc srt: Short report of the silo calculations.
prep:
Preparation report.
calc:
Calculation report of the current method.
act dir:
Methods of the current directory of the memory card.
mem card: All methods of the memory card.
all:
All reports.
ff:
Form feed on printer.
Reports which are sent from the Titrino are marked with space (ASCII 32) and
' at the beginning. Then an individual identifier for each report follows. Reports
which are triggered by RS232 ($G) have the same introducer but without pre-
ceding space, i.e. they start with '.
3.2.2.131. Info.CalibrationData
Info.CalibrationData.Inp1.pHas
Info.CalibrationData.Inp1.Slope
Info.CalibrationData.Inp1.Temp
Info.CalibrationData.Inp1.Date
Info.CalibrationData.Inp1.ElectrodeId
identical for .Inp2 and .Diff
pH calibration data for measuring input 1. After the calibration, the data are
entered automatically together with the date of the calibration and the elec-
trode identification, see 3.2.2.71.
scalc srt, prep, calc, act dir, mem card, all, ff
-20.00...7.00...+20.00
-9.999...1.000...+9.999
-170.0...25.0...+500.0
$G
read only
read only
799 GPT Titrino