799 GPT Titrino
3.2.2.28. Mode.Parameter.Presel.ActPuls
Output of a pulse on the I/O line "Activate", see page 224.
3.2.2.29. Mode.Parameter.SET1.EP
Mode.Parameter.SET1.UnitEp
Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and
Ipol) resp. uA (with Upol). The corresponding unit can be read with .UnitEP.
If the value is on "OFF", no further nodes will appear from SET1.
3.2.2.30. Mode.Parameter.SET1.Dyn
Mode.Parameter.SET1.UnitDyn
Mode.Parameter.SET1.MaxRate
Mode.Parameter.SET1.MinRate
Parameters for SET: Control parameters, see page 31.
.Dyn:
Dynamics, control range in pH, mV (with U and Ipol) or uA (with
Upol). The corresponding unit can be read with .UnitDyn.
.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum
possible rate with the Exchange Unit in current use.
.MinRate: Minimum titration rate in ul/min.
3.2.2.31. Mode.Parameter.SET1.Stop.Type
Mode.Parameter.SET1.Stop.Drift
Mode.Parameter.SET1.Stop.Time
Mode.Parameter.SET1.Stop.StopT
Parameters for SET: Type and size of the stop criterion of the titration.
.Type: Type of stop criterion after stop drift or switch-off delay time.
.Drift: Stop drift in ul/min. Applies when "drift" has been selected.
.Time: Switch-off delay time in s. Applies when "time" has been selected. "inf"
means infinite.
.StopT: Stop time in s. Applies when "time" has been selected and the value of
.Time is set to "inf.".
3.2.2.32. Mode.Parameter.TitrPara.Direction
Parameters for SET, KFT, STAT: Titration direction.
"auto" means the titration direction is determined automatically by the instru-
ment. If 2 EP's have been set in a SET titration, the titration direction is given
by the two EP's. The entry of the titration direction is then invalid.
3.2.2.33. Mode.Parameter.TitrPara.XPause
Parameter for SET, KFT: Pause time in s. Runs before dosing the start vol-
ume.
3.2 Remote control commands
for SET, KFT: first, all, cond., OFF
pH:
0...±20.00, OFF
U, Ipol:
0...±2000, OFF
Upol:
0...±200.0, OFF
read only
pH: 0.01...20.00, OFF
U, Ipol: 1...2000, OFF
Upol: 0.1...200.0, OFF
0.01...10...150, max.
0.01...25.0...999.9
0...10...999, inf
0...99 999, OFF
ON, OFF
read only
drift, time
1...20...999
+, -, auto
0...999 999
149