Section 6: SPS Ultrasonic Scaler Description
6.1 System Controls
Top Grooves
Provide a convenient, temporary
palcement for the handpiece during
treatment procedures. Always place
handpiece in holder when not being used.
Power Adjustment
Turn knob to select ultrasonic
power level for operation:
clockwise increases system
power, counter clockwise
decreases system power.
AC Power Indicator Light
Lighted when System is
switched ON.
Power Control
ON/OFF switch located at the
left front underside of the
Sytem.
6.2 Handpiece
The handpiece accepts all Cavitron
The open end of the handpiece accepts all Cavitron
Insert Port
ENGLISH • 6
Handpiece Holder
Securely holds the System's Handpiece
when system is not being used.
®
30K
™
Ultrasonic Inserts.
30K inserts.
®
Steri-Mate™
Handpiece
Lavage Flow Adjustment
Turn knob to select Lavage Flow level during System operation.
Clockwise increases flow at insert tip, counter clockwise decreases
flow.
The rate of flow through the Handpiece determines the temperature
of the lavage. Low flow rates produce warmer temperatures, higher
flow rates produce cooler temperatures.
Handpiece Connector-
(Handpiece & Mating
Assembly are Keyed)
Date Codes (MMY)
Prior to connecting, align Handpiece and
Cable Assembly electrical connections. If
Cable Assembly does not seat into Handpiece,
gently rotate handpiece until contacts align.
Handpiece
Operates all Cavitron
®
30K™
Ultrasonic inserts and trans-
mits power and lavage from
the System to the insert.
Boost Indicator Light
The two-position Foot Control
provides the capability to
increase the front panel Power
dial setting. When lighted,
indicates Foot Control is fully
depressed, and System is
functioning in"Boost" mode.
Cable Assembly