For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic immunity
The 04-825-001 is intended for use in the electromagnetic environment specified
below. The customer or the user of the 04-825-001 should assure that it is used in
such an environment
IEC 60601
IMMUNITY test
test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
04-825-001 IM_13Feb2015.indd 26-27
Table 2
Electromagnetic
Compliance level
environment
guidance.
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
± 6 kV contact
with synthetic
± 8 kV air
material, the
relative humidity
should be at least
30 %.
Power frequency
magnetic fields
should be at levels
characteristic of a
3 A/m
typical location in a
typical commercial
or hospital
environment.
26
Table 3
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic immunity
The 04-825-001 is intended for use in the electromagnetic environment specified
below. The customer or the user of the 04-825-001 should assure that it is used in
such an environment.
IMMUNITY test
IEC 60601test
Compliance level
Radiated RF
3 V/m 80 MHz
IEC
3 V/m
to 2.5 GHz
61000-4-3
Electromagnetic
environment guidance.
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the 04-825-001,
including cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
27