v The test does not start after applying the
sample.
Probable Cause
Defective test strip.
Sample applied after
automatic shut-off.
(Three (3) minutes
after user action)
Defective meter.
AutoCode
®
Actions
Repeat the test with a new
test strip.
Repeat the test with a new
test strip.
Apply sample only when
the " " symbol appears on
the display.
Please contact Customer Care
at 1.800.243.2636 for assistance.
Owner's Manual
Blood Glucose Meter
59