430-11902-02 Rev N
Notes:
1. Power and impedance measurements are specified at the ESU output
connections and do not account for effects that the cable or the probe may
have.
2. Source Current, Patient Leads, All Outputs Tied Together
ENVIRONMENTAL SPECIFICATIONS
Table 14-2 lists the environmental specifications for the System.
Table 14-2: Environmental Specifications
Description
Storage/Transport
Temperature
Relative Humidity
Operation
Temperature
Relative humidity
Altitude
LABELING SYMBOLS
Table 14-3: Labeling Symbols
Image
Specification
–40°C to 70°C (-40°F to 158°F)
The unit should be gradually returned to
the operating temperature range before
use and stabilized for one hour before
operation.
10% to 90%, non-condensing
The unit should be gradually returned to
the operating environmental conditions
before use and stabilized for one hour
before operation.
10°C to 40°C (50°F to 104°F)
30% to 75%, non-condensing
0 to 2000 meters
Description
Caution
Dangerous Voltage
Warning, Electricity
Defibrillator-Proof Type CF Equipment
Page 35 of 411