4
M3752A AED TRAINER 2
F
Up/Down Buttons. Control scenario selection, adjust voice prompt
volume during use. IMPORTANT: These buttons have different functions on
the real automated external defibrillators. Please refer to the FR2+
Instructions for Use.
G Status Window. This window simulates the FR2+ status display. By
pulling on the tab you can switch between the hourglass symbol (ready
for use) and the red X (troubleshooting required).
H Battery. The AED Trainer 2 battery resembles that of the FR2+
Defibrillator. Insertion and removal mechanisms are similar. However,
the AED Trainer 2 battery case holds six (6) replaceable C-cell alkaline
batteries and can only be used with the AED Trainer 2.
I
Training Pads. Training pads resemble those
used with the FR2+ Defibrillator, however
they do not deliver a shock. Training pads
with a yellow connector are for use with any
manikin; those with a white connector are
for use with the AED Little Anne or Resusci
Anne manikins with Laerdal Link Technology.
NOTE: The AED Trainer 2 comes with one set of training pads with a yellow
connector. When the yellow connector is inserted into the trainer's connector
socket, the trainer responds as if the pads are attached to a patient and
proceeds through the treatment protocol. The training pads are for training
use only and will not function with an actual FR2+ Defibrillator.
J
Remote Control. The optional remote control can be used to control
the different features of the AED Trainer 2 during training sessions.
(See the table on the next page for descriptions of individual functions.
Most remote control actions generate a response in the operator
display window on the AED Trainer 2 device as shown in the table.)