4.4
Configuration of the SCE-020-03
In this version the input signals are frequency (50 mV
than analog signals. In contrast to the named variants SCE-020-01
and –02, there is no analog output available, so that a setting of the
measurement range is inapplicable. The parameters to be set
consist of the decimal place (analogous to programming the
SCE-020-01, see Sections 4.1 and 4.2) and the parameter FACt (K-
factor) or iPr (impulse/rotation). The menu is set up according to this
parameter configuration.
The configuration menu is activated through the
with the menu
, the desired menu is selected with the +/- keys:
End
End, ...., FACt, iPr.
Fig.11:
Menu-assisted operations SCE-020-03
The decimal place must be configured according to Section 4.2.1
first !
Depending on the measurement task, different configurations result
from the menu. The most important applications are as follows:
.. 5 V
ss
SET
- 18 -
) rather
ss
key. Starting