RUHOF ENDOZIMER® AW PLUS WITH APA at dilution ratio: 1:270,
Temperature: ˂60°C and contact time: 5 mins). C4) Rinse further with
pure water at room temperature of 25 °C (or distilled water/deionised
water) for twice with each rinse lasting for at least 1 minute. C5) In the
washing stage, the water temperature should not exceed 45 °C, otherwise
the protein will solidify, and it is difficult to remove.
d) Disinfection:(d1)Use high temperature disinfection function. Set the
temperature to 93°C and disinfect it under the temperature for 5 minutes.
(d2)Start disinfection.
e) Only distilled or deionized water with a small amount of
microorganisms (<10 cfu/ml) can be used for all rinsing steps.
(For example, pure water that is in accordance with the European
Pharmacopoeia).
f) After cleaning, the chemical residue should be less than 10mg / L.
g)The air used for drying must be filtered by HEPA.
h) Regularly repair and inspect the disinfector.
6.5 Drying
If your cleaning and disinfection process does not have an automatic
drying function, dry it after cleaning and disinfection.
Methods
1. Spread a clean white paper (lint free white cloth) on the flat table,
point the product against the white paper (white cloth), and then dry
the product with filtered dry compressed air (maximum pressure 3 bar).
Until no liquid is sprayed onto the white paper (white cloth), the product
drying is completed.