Cleaning, Disinfection And Sterilization - B.A. International OPTIMA 10 BASES10 Modo De Empleo

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 26
affect the light intensity.
5.8 This equipment will turn off automatically if no any action within 2
minutes, turn it on by press any button.
5.9 The depth of solidification of composite is no less than 4mm per 10
seconds.
5.10 The optical fiber should be sterilized for 4 minutes with 134 ℃ and
2.0bar~2.3bar (0.20MPa~0.23MPa) before each use.
5.11 The curing light is equipped with over-heat protection system. It can
continuously work 200s, For example, continuously operate the curing
light for 10 times under 20s working mode (even the curing light works
less than 20s, it is counted as a full operation), then it will come into
over-heat protection status. And only after 2-minute sleep, it can restart
working 200s continuously.

6. Cleaning, Disinfection and Sterilization

The cleaning, disinfection and sterilization of optical fiber is as follow.
Unless otherwise stated, they will be hereinafter referred to as "products".
Warnings
The use of strong detergent and disinfectant (alkaline pH>9 or acid
pH <5) will reduce the life span of products. And in such cases, the
manufacturer takes no responsibility.
This device shall not be exposed to high temperature above 138℃.
Processing limit
5

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido