Electromagnetic Immunity
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the ME Equipment is
used exceeds the applicable RF compliance level above, the ME Equipment should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the ME Equipment.
1
U
: Rated voltage for the equipment.
T
Test Specifications for ENCLOSURE PORT IMMUNITY to RF
Wireless Communication Equipment
Test
Band (a)
Frequency
(MHz)
(MHz)
385
380-395
450
430-470
710
745
704-787
780
810
870
800-960
930
1 720
1 700-1
Service (a)
TETRA 400
GMRS 460, FRS 460
LTE Band 13, 17
GSM 800/900, TETRA
800, iDEN 820,
CDMA 850, LTE Band
5
GSM 1800; CDMA
Modulation
Maximum
(b)
Power
(W)
Pulse
modulation
1,8
(b)
18 Hz
FM (c)
+/- 5 kHz
2
deviation
1 kHz sine
Pulse
modulation
0,2
(b)
217 Hz
Pulse
modulation
2
(b)
18 Hz
2
Pulse
63
Distance
Immunity
(m)
Test Level
(V/m)
0,3
27
0,3
28
0,3
9
0,3
28
0,3
28